Ultrasound in Infertility & Gynecology: Text & Atlas Chaitanya Nagori, Sonal Panchal
INDEX
×
Chapter Notes

Save Clear


Transvaginal ScanCHAPTER 1

 
INTRODUCTION
Evaluation of the female genital tract is the chief diagnostic modality for any gynecological disease and basis of success of any assisted reproductive technology (ART). Ultrasound (US) is the most convenient and reliable modality for evaluation of uterus, fallopian tubes and ovaries. Though these pelvic organs can be assessed by transabdominal, transrectal and transvaginal US; transvaginal route gives best visibility and accuracy of diagnosis.
This is so because
  • Transvaginal probe can be placed close to these organs-uterus, tubes and ovaries.
  • Transvaginal probe is a high frequency probe (6–12 MHz) and therefore has better resolution.
  • The additional advantage is that the patient does not have to tolerate the inconvenience of full bladder.
But transvaginal route cannot be used in virgins or in patients with local vaginal problems, and in these patients transabdominal or transrectal route should be preferred. Transabdominal route has a disadvantage of poor resolution because of more distance of the pelvic organs from the probe, maternal fat, maternal bowel loops and the low frequency (3–5 MHz) probe which has inherently lower resolution than the high resolution, high frequency transvaginal probe. Approximately 42% of ovarian details are missed by transabdominal scan.1
Transrectal sonography in such cases is more preferable and informative as compared to transabdominal approach. The probe used is transvaginal or endocavitary probe, (Figures 1A and B) a high resolution probe so resolution is very similar to that of a transvaginal examination. But the disadvantage is that rectal placement is much more painful than transvaginal placement and needs bowel preparation. Moreover the orientation of pelvic organs that appear on the image created by transrectal scan is different than that of a transvaginal scan and so interpretation may be a little difficult.
Though we shall discuss here in detail the scanning methods of all three routes, we shall discuss transvaginal scan first as this is the most commonly used one.
2
zoom view
Figure 1a: Endocavity (transvaginal) probe
zoom view
Figure 1b: Endocavity (transvaginal) volume probe
 
TRANSVAGINAL SCAN
 
Method
At least a verbal consent of the patient is essential. Patient is asked to empty the bladder. Patient is undressed, placed in lithotomy like position (Figure 2) on the gynecology couch in the same way as for per speculum or per vaginal examination and is adequately covered with a clean sheet, so that she does not feel uncomfortable. Maintaining privacy and dignity of the patient is of utmost importance. Gynecology couch, as it has a gap in the center, allows easy probe movements (Figure 3). A pillow under the patient's buttocks may help, to raise the buttocks if the scan is done on flat bed. Ultrasound jelly is put on the head of the transvaginal probe and then the probe is covered with the condom, not leaving any air between the probe and the condom.
zoom view
Figure 2: Diagrammatic presentation of the patient position and preparation for transvaginal scan
zoom view
Figure 3: To allow up and down movement of the probe as shown in the diagram, the gap of the gynecology couch is convenient
A small amount of jelly is then placed over the condom on the probe head and the probe is gently slided into the patient's vagina. (Figure 4). Counseling the patient before examination and explaining the whole procedure, helps eliminate the anxiety and resistance.
3
zoom view
Figure 4: Position of the probe position in the vagina
 
Tips for Probe Insertion
In case of difficulty in introduction or patient's resistance to introduction, she is advised to take deep long breaths with open mouth, i.e. deep inspirations and long complete expirations. In spite of that if introduction of probe is difficult, the pressure should be exerted posteriorly towards the rectum which will make introduction of the probe into the vagina easier.
 
Basic Probe Movements
There are basic four types of probe movements used for transvaginal scan.
  1. In and out (Figure 5)
  2. Side to side or spanning movement (Figure 6)
  3. Rotation movement or screwing movement (Figure 7)
    4
    zoom view
    Figure 5: Arrow showing in and out movement direction of the probe
    zoom view
    Figure 6: Arrow showing side to side movement (spanning movement) of the probe
    zoom view
    Figure 7: Arrow showing rotation movement of the probe
  4. Up and down or anteroposterior movement (in relation to the patient) (Figure 8).
1. In and out movement
It is used at the probe entry and exit from the vagina. This movement helps to slide the tissue planes over one another. (Video 1). It is used to diagnose adhesions between uterus and ovaries, ovaries or uterus with bowels loops or anterior and posterior vaginal walls with bladder wall or rectal wall.
2. Side to side movement or spanning movement
This is moving the probe from right side of the patient to the left side of the patient or vice versa, without rotation.
zoom view
Figure 8: Arrow showing up and down movement of the probe
But spanning may be done with the probe in rotated in any direction (Video 2). It is used for the survey of anatomy of the pelvis.
3. Rotation movement or screwing movement
Rotation movement of the probe does not include movement of the probe position. It is only clockwise or anticlockwise rotation of the probe. This movement is used to visualize a particular organ or lesion in different axis. Rotation of the probe can be done in clockwise or anticlockwise direction (Video 3).
4. Up and down movement
This is only angulating the probe towards roof or floor, i.e. is patient's anterior or posterior aspect. Moving the handle up, directs the probe head down (posterior aspect of patient) and moving the handle down, directs the probe head up (towards patient's anterior aspect). This movement along with the spanning movement is used for survey of the anatomy and to find out the position of structure or an organ (Video 4 probe movement).
 
Equipment Settings and Optimizing the Image
Usually, there are presets on all ultrasound equipments for different scans and there is one for each scan and there is one for gynecological 5scans also. The advantage of using a good and proper preset is that one has to manipulate the equipment settings minimally and still get good quality images with majority of the scans. A good quality B mode image is a primary requirement for good quality color Doppler, volume ultrasound images and over and above all for correct diagnosis. Though most of the times, the presets work well, several parameters need to be adjusted with each patient, due to variations in physique, habitus and anatomy. Even if the physics part of it is omitted, it is essential to know about the knobs, what do they do and when to use them.
The knobs that need to be often manipulated during each scan to get optimum information out of the scans are scanning angle, scanning depth, probe frequency, focal zone, Zoom, gains, contrast and probe power.
 
Scanning Angle
Each probe has a maximum scanning angle. This is the maximum angle upto which the ultrasound beam can fan out. It indicates as to how much area can be covered by a single ultrasound beam or how much sidewards can an ultrasound beam see. Maximum scanning angle for transvaginal probes usually vary from 80°–180°. Large scanning angle is very convenient for obtaining the bird's eye view of the pelvis, (Figure 9) but it decreases the speed of scanning which is indicated by frame rate.
zoom view
Figure 9: Wide scanning angle, showing a wide image but a low frame rate indicted by Hz on the top of the screen by arrow
It is known that the real time B mode images that we are seeing on the scanner screen appear continuous because of several static images seen quickly one after the other. Faster the change in frames more real time it looks. Frame rate is the number of static images captured in a unit time by the scanner. This means that if the frame rate is low the image resolution is low, it appears less clear and reveals fewer details. Therefore, once the area of interest has been located, the scanning angle should to narrowed down to just a little larger than the area of interest. Narrowing the scanning angle frame rate, and therefore results in better resolution image (Figure 10).
 
Scanning Depth
Each probe has a limit of maximum depth upto which the ultrasound beam can penetrate. This of course depends on frequency of the sound wave. Low frequency sound wave can penetrate deeper and high frequency sound wave can penetrate shallow depths. Maximum achievable depth by any probe may be used for the initial survey, but then depending on the depth at which the area of interest is, the scanning depth can be decreased. This maneuver again increases the frame rate and therefore improves resolution (Figures 11A and B).
zoom view
Figure 10: Narrow scanning angle, showing a narrow image but a high frame rate indicted by Hz on the top of the screen by arrow
6
zoom view
Figures 11A and B: (A) B-mode ultrasound image of the uterus where in uterus is incorporated only in upper 1/3rd of the image. The lower 2/3rd of the image shows bowel loops and other pelvic tissues which are not of importance for the diagnosis of the uterus. This means that the total ultrasound energy is used to scan a deep area (8 cm depth is marked on the top of the screen by white arrow and frame rate by yellow arrow); (B) B-mode ultrasound image of the uterus where in uterus is seen in > 2/3rd of the image. This means that the total ultrasound energy is used to scan an area with limited depth (4 cm depth is marked on the top of the screen by white arrow and frame rate by yellow arrow). This evidently improves frame rate and the image quality, revealing more anatomical details
zoom view
Figures 12A and B: (A) B-mode image of the ovary showing follicles, old corpus luteum (white arrow) and stromal details. Vessel deep to the ovary is not well defined. The circle on the top of the image shows a scan depth of 6 cm, scanned with normal setting (higher frequency of the probe); (B) The same probe position, scanned with penetration setting—lower frequency, depth of 9 cm is scanned. The ovarian details are not as clear as in the previous image but the vessels deep to the ovary are much better seen. The images show that higher frequency gives better visibility for superficial structures and lower frequency gives better visibility for deeper structures
 
Probe Frequency
As discussed earlier lower frequency ultrasound has higher penetration capacity, but has lower resolution, whereas higher frequency sound beam has less lower depth penetration but higher resolution. The transvaginal probes available now are multifrequency probes and one can select the frequency depending on the depth penetration required. The options present on the scanner may be high, medium or low frequency or adipose, normal and penetration setting (Figures 12A and B). Commonly available probes are 5–9 MHz and 6–12 MHz.
7
zoom view
Figure 13: Arrow head on the figure on left side is much higher than the level of the endometrium and the endometrial delineation is very poor. Whereas in the image on the right side only focal zone is shift down to the level of endometrium, no other changes are made in the settings and that makes the endometrium much better. Focal zone should be set at the level of area of interest
 
Setting the Focal Zone
The ultrasound beam behaves like a light wave passing through the convex lens and therefore concentrates at a plane called focal zone. Focal zone is the level at which the image is the sharpest. Therefore, focal zone is always set at the level of area of interest. The arrow head on right side of the image indicates the focal zone. There is also an option of having more than one focal zone, when there are multiple levels at which the image needs to be sharp. But increasing the number of focal zones, decreases the frame rate and therefore usually single focal zone is selected (Figure 13).
 
Zoom
After the overview scanning and decreasing the scanning angle and depth, further improvement in the image quality can be achieved by zooming the image. One may zoom the whole image (Pan-zoom) or may use a zooming box to decide and define which part of the image needs zooming (HD zoom). Panzoom only enlarges the image and therefore increases the distance between the image pixels and deteriorates the image quality (Figures 14A and B).
zoom view
Figures 14A and B: (A) unzoomed B-mode ultrasound image of the ovary; (B) Pan-zoomed B-mode mode image of the ovary, showing a larger image, does help to appreciate the details a little better, but does not change the frame rate
Whereas HD zoom concentrates pixels into a smaller box and thus improves image resolution. Larger image can depict more anatomical details and also results in smaller human errors when measurements are done (Figure 15). Image should be zoomed large enough to fill up at least 3/4th of the screen.
 
Gains
The ultrasound wave when emitted from the probe into the body, it hits several tissue planes on its onwards as well as return path. With each plane that it crosses, i.e. that is with the distance that it traverses, it attenuates. This leads to darker (hypoechoic) image of the structures far from the probe than what they are and this may 8cause an error of diagnosis.
zoom view
Figure 15: HD zoomed image of the ovary shows much more details in the ovary, with marked increase in frame rate. The small image in the box in lower left corner, shows a box with yellow outine, which identifies the area to be zoomed
A solid structure may appear cystic. Image can be made brighter by increasing the gains. This is strengthening of the returning beam to make the image brighter. But changing gains inadvertently may lead to erroneous diagnosis. Therefore in all doubtful, difficult situations, the gains are set in such a way that the urinary bladder appears anechoic and echogenicity of other structures are assessed in relation to bladder. (Figures 16A and B) (Video 5). Gains can be adjusted in two ways. Total gains can be increased/decreased or gains can be adjusted layer wise, depending on the distance of the tissue plane from the probe. The later one is known as TGC (time gain compensation). Ultrasound wave returning from deeper structures takes longer time to return and is attenuated. This control compensates for the gains lost by time. (Video 6)
 
Probe Power
Brightness of the image can also be adjusted by increasing the power of the incident beam. Gain adjustment corrects the returning beam and power affects the incident beam. Probe power can be increased maximum to 100%, but usually set at between 80–90%. This decreases the total mechanical and thermal energy transmitted to the tissues. (Figure 17).
zoom view
Figures 16A and B: (A) B-mode ultrasound image of ovary showing corpus luteum(cystic structure with internal echogenecities) on left side. The right half of the image because of low gains, obscures the internal echogenecities and may misdiagnose the structure as a follicle instead of a corpus luteum; (B) B-mode ultrasound image of ovary showing two follicles (cystic structure with no internal echogenecities) on left side. The right half of the image because of high gains, shows internal echogenecities and may misdiagnose the structure as a corpus luteum instead of follicle
 
Contrast
Increased contrast means more black and white image with less shades of gray, and less contrast means more shades of gray in an image. High contrast (Figure 18A) may therefore hide the details of soft tissues and so is usually selected for study of cystic structures or bones. Low contrast (Figure 18B) may mask calcifications. Contrast setting (Figure 18C) may also be presented as dynamic contrast on the ultrasound machine.
9
zoom view
Figure 17: Switch board of the ultrasound scanner shows the icons (arrow), to identify the knob that is used to change the power of the probe
 
Systematic Scan
Probe is introduced in the longitudinal position. That is the position in which the indicators (arrow) on the probe remain up, facing the roof of the room, that is on anterior aspect of the patient (Figure 19). These indicators on the probe match the indicator (logo of the company-circle on the screen (Figure 20). This means that if the indicator on the screen is on the right side of the screen, the structures anteriorly placed are on the right side of the screen and vice a versa. Uterus is normally seen in the midline with this probe position. If uterus is not seen, in the center, the probe is spanned from one side in the pelvis to the other side to search for the uterus. Once the uterus is located with the probe in the same position (without rotation), it is spanned from right to left side of the patient, scanning the uterus in long axis from one side to other (Video 7).
Now rotate the probe 90° anticlockwise. This maneuver will give transverse view of the uterus (Figure 21) with the right side of the patient on right side of the screen. In this transverse position also the whole uterus is evaluated from the fundus to the cervix by sliding the probe up and down in the vagina. (Video 8). Now span the probe towards right side and in transverse position only at the level of uterine cornu, gradually look at the adnexa and follow it to the ovary (Video 9). Extend the movement upto the lateral pelvic wall if ovary is not located.
zoom view
Figures 18A to C: (A) B-mode ultrasound image of the uterus with low contrast, where the differentiaton between different shades of gray is difficult (low contrast); (B) B-mode ultrasound image of the uterus with high contrast, where the image chiefly shows only two colors black and white with no shades of gray (high contrast); (C) B-mode ultrasound image of the uterus with optimum contrast. Both, high and low contrast obscure anatomical details
Locate and assess the ovary in a true transverse section and then rotate it 90° 10to get a true long axis of the ovary, the largest longitudinal diameter (Video 10).
zoom view
Figure 19: Transvaginal volume ultrasound probe showing pits for fixation of biopsy guide, shown by arrow, which act as indicators to define the position of the probe, which is indicated on ultrasound image
Assess the ovaries for any pathology (Video 11). Then span the probe, come back to the midline and follow the same procedure for the opposite adnexa. In and out movement of the probe with the ovary in view or pressing from the abdomen when probe is looking at the ovary and checking the sliding of the organs against each other confirms mobility of the ovaries and rules out adhesions (Videos 12 and 13). This is also known as Timor-Trsitch sign or a sliding organ sign.
zoom view
Figure 20: By the company logo (shown by arrow) on this longitudinal section off the uterus
zoom view
Figure 21: Anticlockwise 90° rotation of the probe from a position in which long axis of uterus is seen, shows transverse section of the uterus on B-mode ultrasound image
Come back to the midline and angulate the probe head posteriorly and look for fluid in pouch of Douglas. While removing the probe gently, have a close look at the cervical canal. Finally slide the probe against the anterior vaginal wall and posterior vaginal wall and check for mobility of tissue planes by sliding sign (Video 14). This completes the 2 dimensional or B-mode routine transvaginal scan (TVS).
Color, pulse or power Doppler can be added as and when required.
Which vessels to interrogate and how to trace them?
For color Doppler of the uterine artery the probe is brought back in the midline in longitudinal plane and then again moving laterally, serpinginous tubular structure is seen at the level of internal os, which is the uterine artery (Video 15) (Figure 22). It can also be traced in transverse axis of the uterus, moving the probe to the level of internal os, when arteries will be seen on both the sides (Figure 23 (arrow)). On abdominal scan, moving from midline laterally at the level of symphysis pubis shows iliac vessels, which are large prominent vessels running mediolaterally, superioinferiorly. After tracing these vessels, the probe if angled slightly medially, shows a vessel perpendicularly crossing them, this is uterine artery (Video 16). Color Doppler is also used to evaluate the endometrial flow. The blood vessels entering the endometrium are branches of spiral artery and are seen as vessels perpendicular to the endometrium (Figure 24).
Moving laterally from the transverse section of uterus towards adnexa will show another vascular structure heading towards the ovary, which is the ovarian vessel (Figure 25). The ovarian stromal vessels are studied for baseline scan . These are the vessels which are lying in the ovarian stroma (Figure 26) and not close to the follicles. Preovulatory scan shows follicles in the ovary with perifollicular flow.
zoom view
Figure 22: Color Doppler and spectral Doppler image of uterine artery (arrow), seen on the long section of the uterus, seen on the lateral most aspect of the uterus
zoom view
Figure 23: Transverse section of the uterus at the level of internal os showing lucent serpinginous vascular structures on both sides (arrows)— uterine arteries
 
DOPPLER TECHNIQUE
 
What is Doppler?
When sound is reflected from a moving object the returning wave has a different frequency than the transmitted frequency, which is known as Doppler shift. The frequency is more than the transmitted frequency when the object moves towards the probe and is less if the object is moving away from the probe.
12
zoom view
Figure 24: Power Doppler showing uterine vasculature with arcuate arteries parallel to the endometrium (small arrow) and spiral arteries perpendicular to the endometrium, penetrating the enometrium (large arrow)
zoom view
Figure 25: Ovarian vessel
Doppler shift (change in frequency) is proportional to the velocity of the object and the incident beam angle and frequency. fD = fr – f0 = 2f (S/v) cos θ.
With constant flow, increase in Doppler angle, decreases frequency of Doppler shift. Lower angle of incidence shows higher frequency shift and therefore shows brighter color and aliasing. Higher incident angle presents lower frequency shift and therefore darker color. For this reason, the angle of the incident beam to the vessel is important. If it is 90° then cosθ is 0 and the results are inaccurate, therefore Doppler angle should always be < 60 °.
zoom view
Figure 26: Color Doppler image of the ovary showing color flow in the central solid part of the ovary (arrow), away from the follicles which indicate stromal vessels
Doppler shift is proportional to opening frequency of transducer. For example, 6 MHz transducer yields double the measured shift than 3 MHz transducer for the same flow. This means, higher frequency increases sensitivity to low flow and lower frequency increase penetration, allows to pick up flow in vessels far from the probe and reduces aliasing.
 
Types of Doppler
The physics and facts of different types of Doppler used for diagnosis is complicated and therefore we shall here discuss only the practically relevant points.
 
Color Doppler
This is a semiquantitative directional Doppler. Flow towards the probe appears red and flow away from the probe appears blue (Figure 27). The brightness of the color represents the intensity of the flow.
 
Power Doppler
This is a nondirectional Doppler. Any moving object represents color (Figure 28).
13
zoom view
Figure 27: Color Doppler of the head showing two middle cerebral vessels, the proximal one carrying blood towards the probe shows red color the distal one carrying blood away from the probe shows blue color
zoom view
Figure 28: Power Doppler showing uterine vascularity. All vessels show same color because power Doppler is non-directional Doppler and also is a better tool to pick up low velocity blood flows
 
Pulsed Wave Doppler/Spectral Doppler (Figure 29)
This is a directional quantitative Doppler. The flow signals are marked as a spectrum. Flow towards the probe is seen above the baseline and flow away from the probe is seen below the baseline. The height of the signal from the baseline on the spectrum can be measured and represents the velocity of the flow.
zoom view
Figure 29: Spectral Doppler of the uterine artery. This is a display of the blood flow pattern on the spectrum, where in flow towards the probe is seen above the baseline and flow away from the probe is seen below the baseline. The amplitude of the flow is seen in the form of height of the spectrum on either side
 
Color and Power Doppler Settings
To derive the correct information about the flows, it is important to set the Doppler parameters. These settings are as follows:
1. Color box size
Same as for B-mode scan, size of color box should be kept to the smallest possible to get highest frame rates (Figures 30A and B).
2. PRF (Pulse repetition frequency)
Higher pulse repetition frequency is to be selected for interrogation of blood flows with high velocity and lower PRF for blood flows with lower velocity (Figures 31A and B). Selecting high PRF for low velocity flows lead to elimination of flow information and selecting low PRF for high velocity flows leads to artifact called aliasing.
3. Wall filter
As is known, Doppler is a frequency shift arising due to movement of the target object and with other variables, its amplitude also depends on velocity of the target object. This means that when moving RBCs (red blood cells) are producing high amplitude signals, vessel wall that moves much slower than 14RBCs also produce low amplitude signals.
zoom view
Figures 30A and B: Power Doppler study of the uterine vascularity, large color box (a) gives lower frame rate (arrow) and smaller color box gives higher frame rate. Higher frame rate is required when doing vascular studies as these are dynamic studies and therefore slower frame rate means intermittent information loss
zoom view
Figures 31A and B: (A) Lower PRF setting (arrow) picks up the endometrial and subendometrial flow, but the flow in the uterine artery shows turbulence in spite of no abnormality in flow; (B) Higher PRF is set which leads to total normal homogenous flow in uterine artery but total obliteration of color signals in the endometrium. This means correct PRF settings are important for getting correct flow pattern information
These signals do not produce any useful information and so should be best filtered out. This function is done by wall filter. It restricts low amplitude signals to improve the quality of signals coming from blood flows. Wall filter is set to lowest for low velocity blood flows like follicular or endometrial flows. If the wall filter is high it would cut down the information of the diastolic flow first and also sometimes the information from low velocity systolic flow (Figures 32A and B). For study of very high velocity blood flows like in fetal echocardiography, the amplitude of movement of heart wall is also high and so wall filter setting is also adjusted to high, otherwise there will be lot of noise and the information of blood flows will be polluted.
4. Gains
Gains, like in B-mode settings must be set correctly for color Doppler also. Too much gains leads to spill of color outside the vessels and too little gains leads to cutting off the information of low velocity flows (Figures 33A to C). In any vessel, the blood flow velocity 15is highest in the center, but is low close to the walls, due to friction with the wall.
zoom view
Figures 32A and B: (A) Wall filter cuts off lower intensity signals. If wall filter is set high, (arrow) it cut off all diastolic and some systolic flow information; (B) Normal wall filter setting, showing uterine perfusion on power Doppler (arrow at wall filter)
It is information about this flow that may be missed by low gains or high wall filter. The image of color Doppler in that case would show gap between vessel wall and color column.
5. Balance
This is an important setting. Even with all above mentioned settings, the color filling is dependent on the brightness of B-mode image. The color appears patchy when the B-mode image is too bright. The limit of brightness beyond which the color will appear patchy is decided by balance. On the left upper corner of the screen, two vertical bands are seen, one in gray and one in color. On the gray band, a green line is seen which indicates the balance.
zoom view
Figures 33A to C: (A) Correct gain settings is when color fills up the complete lumen of the blood vessels under interrogation, and no color spills out of the vessels; (B) Low gain settings when blood vessels do not show color filling; (C) High gain setting shows all the color spilling out of the blood vessels
If the brightness of the image is more than the brightness above the green line, the color filling will appear patchy. To correct this either the gains on B-mode are decreased or the 16balance is to be increased. Generally therefore, if this manipulation with every patient is to be avoided, the balance is to be set in such a way that the B-mode image does not become too dark and the color fill up is smooth (Figures 34A and B).
 
Spectral Doppler Settings
1. Sample volume
The sample volume should be equal to the diameter of the vessel. If large may lead to overlapping of spectrum from multiple vessels and if small may evaluate flow only in part of the vessel lumen (Figures 35A and B).
2. Sample angle
This is a very important setting for velocity assessments. It has been described earlier in the equation for Doppler shift, that the shift is dependent on cos θ where θ is the Doppler angle. Because value of cos 0° is one and that of cos 90° is 0, the correct value of velocities can be achieved when the Doppler angle is the smallest. As the Doppler angle increases the error in velocity calculation increases, till it becomes unacceptable beyond 60°. Therefore, Doppler angle should be set at the lowest possible. When spectral Doppler is switched on, a dotted line appears appears on the screen, on which a sample volume appears and also a line that can be rotated. The angle formed between this line and the dotted line indicates the Doppler angle. This line should be adjusted in such a way that it remains as parallel to the vessel as possible and the Doppler angle is the smallest possible. (Figures 36A to C).
3. Gains
Gains should be adjusted in such a way that the spectrum appears clear—only the information of the flow is seen and no information of the flow is being eliminated from the spectrum. Too high gains create noise (unwanted movement information), too low gains eliminate the signals from low-diastolic flow (Figures 37A to C).
zoom view
Figures 34A and B: Balance is deciding a level of brightness on B-mode image. The color pick up will be smooth when the brightness of B-mode image is less than this level and color pick up is patchy when brightness of B-mode image is more than the defined level of brightness. The solution to this problem is either increasing the balance (a) or decrease the gains on B-mode image (b) Any of these two procedures can improve the color pick up
PRF and wall filter settings are same as that for color Doppler (Figures 38A to C) and (Figures 39A and B).
The commonest artifacts seen on Doppler are aliasing and mirror image. Aliasing on color Doppler appears as mixing of colors (Figures 40A and B). This is due to selection of wrong PRF. On spectral Doppler this is seen as overlapping of systolic and diastolic flows or overlapping of the high systolic flow velocities on the other side of the baseline. This can be corrected by either selection of correct PRF or by shifting the baseline.
17
zoom view
Figures 35A and B: (A) Color with spectral Doppler image of a vessel showing arterial trace above the baseline on the spectrum and the sample volume (distance between two horizontal lines lying perpendicular to the long dotted line) is just large enough to include the diameter of the vessel; (B) A larger sample volume includes both artery and the vein and so give a trace on the spectral Doppler on both sides of the baseline—arterial above and venous below
zoom view
Figures 36A to C: After placing the sample volume on the vessel, sample angle is adjusted. when angle is manipulated, a short line peeps out from the dotted line on both sides of sample volume. This line is preferably adjusted as parallel as possible to the vessel to be interrogated. When this rotation is done to adjust the angle, there is an angle created between this short line and the dotted line and this angle is known as sample angle and is displayed on the left side of the image (arrow). Change in this angle changes the velocity readings and this angle should be as small as possible, but always less than 60°. A, B, C respectively show angle of 30°, 22° and 68° and notice the changes in velocity readings
Mirror image artifact 18(Figure 41) is seen on spectral Doppler and can be avoided by selecting a correct sample volume to avoid interrogating two loops of a vessel or two vessels at a time. Other instant when a mirror image artifact is seen is when a large sample volume is placed on the curved vessel.
zoom view
Figures 37A to C: (A) Color and spectral Doppler of uterine artery is seen on a with normal gain settings. (B) With high gains shows the complete spectrum filled with low level echoes (noise), corrupting the actual spectrum, (C) Very low gains of the spectral Doppler shows loss of information on both systolic an diastolic flow
Tips for good Doppler settings:
  • Appropriate sample volume for vessel size.
  • Appropriate PRF for the vessel to be sampled.
  • Sample in the center of the large vessels as close to the walls of the vessels, the flow is slow.
  • Choose proper wall filters—high for large vessels/heart, low for smaller vessels/low flow.
  • Never use a Doppler angle higher than 60°.
  • Make lowest possible sensitivity settings.
  • Use continuous wave for very high velocities.
 
3D and 3D Power Doppler Techniques in Gynecology and Infertility
It is out of the scope of this book to explain the basics of 3D ultrasound and all the applications of volume ultrasound as such. We shall therefore discuss only the relevant applications of volume ultrasound here. Volume ultrasound combined with transvaginal scan has revolutionized the gynecological imaging. Multiple planes and sections available on the tissue blocks of volume ultrasound has made the understanding of the complex anatomy easy. The availability of coronal section is of prime importance. Volume ultrasound has also made volume calculations more accurate.
19
zoom view
Figures 38A to C: (A) PRF (pulse repetition frequency) like in color Doppler is adjusted according to the velocity of the blood flow velocity in the vessel to be interrogated. Correct PRF setting should show the systolic and diastolic flow contained to individual side of the baseline. (B) Too low PRF setting for a high velocity flow shows overlapping of the spectrum from both sides of the baseline. (C) Too high PRF will not allow differentiation between systolic and diastolic flows especially in low resistance flow spectrum
zoom view
Figures 39A and B: Wall filter filters out low intensity Doppler signals, thus not allowing the spectrum to be corrupted by the signals arising from movement of the vessel wall and transmitted pulsations from surrounding tissues. (A) Shows optimum wall filter with no loss of information on the spectrum, whereas (B) Shows high wall filter, cutting the flow signals with lowest intensity and thus cutting down the diastolic flow signals and a black line between the baseline and the spectrum
20
zoom view
Figures 40A and B: Show aliasing on color Doppler and spectral Doppler. This occurs actually when flow is turbulent but when PRF settings are too low for the flow velocity it leads to mixing of colors on color Doppler and over lapping of systolic and diastolic spectrum on spectral Doppler. This is an artefact and not a turbulent flow actually. This can be confirmed by increasing the PRF and repeating the spectral and color Doppler
zoom view
Figure 41: When sample volume is large enough to interrogate two arteries or two loops of the same artery lying close to each other with flow in opposite direction, at a time or the sample volume is exact on the turning point of a curved vessels, a mirror image artefact occurs. Almost equal blood flow is seen on both sides of the baseline
This is of importance for gynecological masses as well as for infertility assessment. New softwares of volume calculation like SonoAVC have great applicabilities in infertility for follicular assessment as well as for other fluid filled masses of fallopian tubes or ovaries and also for volumetric assessment of early pregnancy. New rendering mode—HD Live is very helpful for assessment of uterus, for HyCoSy and for assessment of fetal development.
 
Acquiring the Volume: (Video 17)
The area of interest is first selected on the B-mode scan. This area should ideally be placed in the center of the monitor screen. The B-mode image should be optimized. The volume box is switched on and is placed over the area of interest. The volume box should be large enough to include the area of interest with at least 5–10 mm of margin on all sides. To select the volume angle, probe is manually angulated up and down or side to side depending on orientation of the image to assess the depth the organ of interest. Quality of the volume is selected depending on whether the structure or organ of interest is steady and stable or has movements. Higher quality takes longer to acquire as it contains more B-mode images in the volume block, whereas lower quality images takes shorter time to acquire. Having adjusted the region/area of interest, angle of acquisition and quality, then acquisition is done. The transducer head takes an automatic sweep and the acquired volume is displayed on the screen as three images in three orthogonal planes—x, y, z axis.
21
 
Multiplanar Imaging
Walking through each image would take through the complete anatomical details in the volume box in individual plane. This is multiplanar imaging. This is the study of sectional planes and makes the understanding of anatomical relationship of various structures clear. A dot is seen in each image and this is the point at which the orthogonal planes intersect. This point is known as the reference point as it represents the same anatomical landmark in all three sections. Coronal section is the most important, as this is the plane which is inaccessible on B-mode ultrasound imaging (Figure 42).
 
Rendering
Viewing the whole tissue block as a whole from any one side is known as rendering (Figure 43). When the image is rendered a box appears on each image. In two of the three images, the box is drawn by three yellow lines and one green line and in the third image, the box has all yellow lines. The resulting rendered image will be similar to the image in this later box. The green line on the box is the viewing line. This line is aligned with the curve of the organ of interest. For gynecological scans and infertility scans almost always up-down rendering direction is used. Rendering can be done in various modes and even combination of various rendering modes can be used viz. Surface modes, transparent modes, inversion, angiomode, etc. (Figures 44A to D). For tissue rendering that is required for uterus and ovaries commonly combination of surface texture or smooth with transparent mode is used. Though rendering can be done for straight as well as curved surfaces, it is difficult to render the uterus or any other surfaces with acute curves. It is in these cases that omniview is used (Figure 45).
zoom view
Figure 42: Multiplanar image acquired by 3D ultrasound. The sections are longitudinal, transverse and coronal sections of the uterus in clockwise manner starting from upper left corner and as seen in the symbol in the center of the image marked by the circle, these are known as sections a,b and c respectively
zoom view
Figure 43: 3D ultrasound acquired image of the uterus showing three sectional planes in black and white and a sepia colored image representing the rendered volume. The sectional images show the box, this is known as rendering box, it selects the part of the acquired volume to be seen on the rendered image. Two out of three boxes show a green line. This line is the viewing line, it can be straight or curved. It acts as a window through which the rendered volume is seen. The green line can be changed to make any side of the box on any of the three sectional planes and that decides the rendering direction. In this it is up-down rendering direction, which is almost always used for all gynecological rendering
For 3D power Doppler angiographies, the acquisition is done with power Doppler and rendering is done in glassbody or angiomode (Figures 46A and B) to 22see the soft tissue details and vascular anatomy together or only blood vessels respectively. Using volume histogram with the volume that is acquired with power Doppler, allows quanitative assessment of the global vascular indices..
On omniview, if a line is drawn on the sectional plane on a static 3D or a live 3D (4D) image, the orthogonal plane of the imaging plane can be immediately reconstructed. Omniview when is used with thick slices improves the contrast markedly and this is known as volume contrast imaging (VCI) (Figures 47A and B). VCI can be used in any plane on static 3D US and on scanning plane (VCI A) and coronal plane on 4D US. VCI A is very useful for visualization of subtle lesions like small polyps, small submuosal fibroids, etc. (Video 18).
Volume calculations have become very accurate on 3D US using VOCAL soft ware. After volume acquisition, any one section is selected and that is rotated 180° in all. This total rotation is done step by step. This angle can be selected from 6°, 12° 15° or 30° steps. At each step, the margins of the organ or lesion of interest are outlined and when complete 180° rotation is over, volume of the organ is displayed on the monitor screen (Figure 48).
Another software that is of great help in infertility practise is SonoAVC. It calculates the volume of all fluid filled structures and color codes these (Figure 49).
zoom view
Figures 44A to D: The volume acquired by 3D/4D ultrasound can be rendered in various modes or combination of any two modes, for example—a few are: (A) Surface mode to see the surfaces, (B) Transparent (maximum) mode to see the internal anatomy, viz. bones (C) Inversion mode fluid filled structures appear echogenic and solid looking (D) Angiomode shows only blood vessels as in angiography
23
zoom view
Figure 45: Volume ultrasound of uterus with omniview. This software allows to seen the coronal section of the said structure in the plane of the line drawn. This can be used on 3D acquired volume as well as on live 4D volume
zoom view
Figures 46A and B: 3D Power Doppler volume rendered in glass body mode (A) which shows the soft tissue details and also the vascular details, and angiomode which shows the vascular details only as in angiography
zoom view
Figures 47A and B: 3D ultrasound of the head with omniview (A) showing the midsagittal view of the head when line for omniview is drawn across the cavum septum Pellucidum on the transthalamic axial plane of the head. The same image when coupled with volume contrast imaging (VCI) shows better contrast and better delineation of corpus callosum and midline structures
24
zoom view
Figure 48: 3D ultrasound of the follicle with VOCAL software used to define and calculate the volume of the follicle as seen on the fourth image
zoom view
Figure 49: 3D volume of the ovary rendered by SonoAVC, which shows individual follicle identified by a different color, also calculates volume of each follicle
zoom view
Figure 50: SonoAVC is also used for counting and measurement of the pre-hCG follicles, especially when there is multifollicular development
It is therefore of great help for calculation of antral follicles or for calculation of follicular diameter or volume or multiple pre-hCG follicles of controlled ovarian hyperstimulation (Figure 50). SonoAVC can also be used for volume calculation of hydrosalpinx, cystic adnexal lesions or early gestational sac (Figures 51A and B).
25
zoom view
Figures 51A and B: SonoAVC is based on inversion mode principle and therefore can be used for assessment of the volume of any cystic structures or lesions. Hydrosalpinx is shown on B-mode in figure A and shown on sonoAVC with volume calculation in the lower right corner in figure B
To summarize—
Optimize B-mode images:
Adjust angle, focal zones, focal depth, contrast (Dynamic range), gains, zoom probe power, probe frequency.
Doppler is optimized by:
Size of color box or sample volume, gains, PRF, wall filter, probe power, Doppler angle.
3D is a value adding modality with:
Multiplanar imaging, omniview, VCI, TUI, rendering, VOCAL and SonoAVC.
REFERENCE
  1. Hull MGR. Polycystic ovarian disease: clinical aspects and prevalence. Res Clin Forums. 1989; 11:21-34.