Vibrant SoundBridge (VSB) preparing FMT
by

The Surgical Manual of Common Otological Implants

by Ing Ping Tang, Philip Rajan Devesahayam, Prepageran Narayanan
About Video

This video showing Vibrant SoundBridge (VSB) preparing FMT. Inspect the FMT clip under the microscope. If the clip is narrow, carefully open it by sliding a straight needle through the clip opening until the approximate diameter of the long process of incus is about 0.5 mm. Advance the FMT into the middle ear using a smooth alligator forceps or a suction tip. Avoid grasping the FMT at its junction to the wire.